3.11 Tilt Drive Status Indicator

The tilt drive status indicator light is lit when the power lateral tilt mechanism is in the
center position of the +/- 25 degree powered lateral tilt range. If the tilt drive indicator is
not illuminated, operate the lateral tilt function using the hand pendant in the appropriate
direction until the indicator is lit. The head end crossbar should be horizontal when the
tilt drive status indicator is lit. If the head end crossbar is not horizontal, the internal
power tilt mechanism is out of synchronization with the crossbar. Refer to the section
on synchronizing the lateral tilt function in the 5892 advanced control base user guide or
the 5803 owner's manual, to reset this function.

3.12 Installing The H-Frames

The following coupling procedure assumes the head end and foot end columns do not
yet have either the Jackson Spinal Surgery Top or the Radiolucent Imaging Top in
place. The coupling procedure for the Jackson Spinal Surgery Top or the Radiolucent
Imaging Top is as follows:
1. Lock table base in position by engaging the (4) caster locks.
2. Install an H-frame to head end crossbar and secure with a T-pin (see Figure 25).
Fully retract the head end crossbar against the head end column.
3. Install an H-frame to foot end crossbar and secure with a T-pin (see Figure 26).
Sliding assembly
HEAD END
Figure 25: H-frame installed - head end column
5892/5803 Advanced Control Base
MIZUHOSI 2009
T-pin
ON
H-frame
19
Crossbar
Drop lock
ON
H-frame
Figure 26: H-frame installed – foot end column
5892/5803 Advanced Control Base
FOOT END
T-pin
NW0504 Rev. D