4.0 IMAGING PROCEDURES

To perform most imaging procedures, attach the Radiolucent Imaging Top according to
the following procedure;
1. Lock the table base in position by engaging the (4) caster locks.
2. Verify that the table controls are locked.
4.1 5890/5891 Modular Table Bases
1. Engage the patient transfer safety locks at both the head and foot end columns (see
section 3.0, Figures 13 and 14).
2. Set the rotational friction control to maximum friction at both the head and foot end
columns (see section 3.0, Figures 13 and 15).
3. Engage the 25° rotation stop (see section 3.0, Figure 16).
4.2 5892/5803 Advanced Control Bases
1. Confirm the rotation safety lock switch is illuminated (see section 3. 0, Figure 24).
2. Confirm rotation lock indicator light is illuminated (see section 3.0, Figure 24).
3. Confirm the tilt drive status indicator is illuminated (see section 3.0, Figure 24).
4.2.1 Attach the Radiolucent Imaging Table Top with table pad (see section 3.0: Table
Top Coupling Procedures), depending on which base is used.
4.2.2 Transfer the patient in a supine position to the Radiolucent Imaging Top. Apply
the patient safety strap and secure.
Figure 27: Patient supine on the Radiolucent Imaging Top with pivoting arm boards
MIZUHOSI 2009
21
NW0504 Rev. D