HEAD END
Drop lock
Figure 21: Head end column with coupled Jackson Spinal Surgery Top
Verify that the T-pin is inserted through the H-frame, passes completely through
the table mounting tube and through the opposite side of the H-frame and that the
T-pin drop lock is in sight and pivots freely.
8. To couple the foot end of the table top, slide the H-frame and crossbar as necessary
to accommodate the length of the table top (see Figure 22). Align the mounting tube
on the table top to the selected hole on the foot end H-frame. Push the T-pin
through the H-frame, completely through the mounting tube on the table top and
through the opposite side of the H-frame. This should be done at the same level as
the mounting hole where the head end was coupled. The T-pin drop lock should be
visible and pivot freely.
Jackson Spinal
Surgery Top
MIZUHOSI 2009
FOOT END
H-frame
T-pin
Figure 22: Foot end column with coupled Jackson Spinal Top
H-frame
T-pin
Jackson Spinal
Surgery Top
Warning:
Drop lock
16