3.5 Installing The H- Frames

The following coupling procedure assumes the head end and foot end columns do not
have either the Jackson Spinal Surgery Top or the Radiolucent Imaging Top in place.
1. Lock the table base in position by engaging the (4) caster locks.
2. Install an H-frame to head end crossbar and secure with a T-pin (see Figures 17 and 18).
Drop lock
T-pin
H-frame
Figure 17: H-frame aligned with mounting stud
and T-pin prepared for insertion
Failure to install the T-pin will result in the patient and table assembly being dropped.
3. Install an H-frame to foot end crossbar and secure with a T-pin. Fully retract the foot
end crossbar by sliding it toward the foot end column (see Figure 19).
T-pin
Drop lock
ON
H-frame
Figure 19: H-frame installed – foot end column (5890/5891 base shown)
MIZUHOSI 2009
Crossbar
Mounting stud
WARNING:
Crossbar
Rotation friction control
HEAD END
T-pin
Figure 18: T-pin insertion into H-frame and
crossbar head end column
(5890/5891 base shown)
T-pin
SLIDE
14
Crossbar
H-frame
NW0504 Rev. D