Cleaning; Washing Procedure; Washing Limitations - Stryker Power-PRO IT Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

The Power-PRO™ IT cot is designed to be power washable. The unit may show some signs of oxidation or discoloration
from continuous washing, however, no degradation of the cot's performance characteristics or functionality will occur
English
due to power washing as long as the proper procedures are followed.
Thoroughly clean the cot once a month. Clean Velcro
disinfectant to evaporate. Appropriate disinfectant for nylon Velcro

WASHING PROCEDURE

• Always remove the battery! Never wash the cot with the battery installed.
• Follow the cleaning solution manufacturer's dilution recommendations exactly.
• The preferred method Stryker Medical recommends for power washing the cot is with the standard hospital
surgical cart washer or hand held wand unit.

WASHING LIMITATIONS

WARNING
When cleaning, use any appropriate personal safety equipment (goggles, respirator, etc.) to avoid the risk of inhaling
contagion. Use of power washing equipment can aerate contamination collected during the use of the cot.
CAUTION
• DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE UNIT.
• Maximum water temperature should not exceed 180°F/82°C.
• Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand held wand is being used to wash the
unit, the pressure nozzle must be kept a minimum of 24 inches (61 cm) from the unit.
• Allow cot to air dry.
• Towel dry all casters and interface points.
• Failure to comply with these instructions may invalidate any/all warranties.
• Always remove the battery before washing the cot.
Return To Table of Contents
1-60

Cleaning

®
AFTER EACH USE. Saturate Velcro
®
should be determined by the service.
6516-109-005 REV B
®
with disinfectant and allow
www.stryker.com

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

6516

Table des Matières