WARNING (CONTINUED)
English
®
• The Drager
Power-PRO™ IT cot. Using this adaptor on any cot other than the model 6516 Power-PRO™ IT cot or using any
unapproved incubators in this configuration may result in damage to the cot and/or injury to the patient or user.
• Stryker is not responsible for specifications changes to the Drager
• The Airborne™ Stackable adaptor (6516-127-000) adaptor is designed to secure only an Airborne Stackable to the
model 6516 Power-PRO™ IT cot. Using this adaptor on any cot other than the model 6516 or using any unapproved
incubators or stackables in this configuration may result in damage to the cot and/or injury to the patient or user.
• The Air Sled, no adaptor option (6516-142-000) is designed to secure incubators without an adaptor to the
model 6516 Power-PRO™ IT cot. Using this configuration on any cot other than the model 6516 Power-PRO™ IT
cot or using any unapproved incubators in this configuration may result in damage to the cot and/or injury to the
patient or user.
• Stryker is not responsible for specification or option changes to Air Sled compatible incubators.
• When the optional head end storage flat is being used, ensure that it does not interfere with the operation of the
retractable head section, safety bar and safety hook. Injury to the patient or operator could result.
• When cleaning, use any appropriate personal safety equipment (goggles, respirator, etc.) to avoid the risk of
inhaling contagion. Use of power washing equipment can aerate contamination collected during the use of the cot.
• SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT
IF USED IMPROPERLY. If the products described above are used to clean Stryker EMS equipment, measures
must be taken to ensure the cots are wiped with clean water and thoroughly dried following cleaning. Failure to
properly rinse and dry the cots will leave a corrosive residue on the surface of the cots, possibly causing premature
corrosion of critical components.
• Failure to properly clean or dispose of contaminated cot components will increase the risk of bloodborne pathogens
and may cause injury to the patient or operator.
• Escaping fluid under pressure can penetrate the skin causing serious injury. Avoid the hazard by relieving pressure
before disconnecting hydraulic or other lines. Tighten all connections before applying pressure. If an accident
occurs, see a doctor immediately. Any fluid injected into the skin must be surgically removed within a few hours or
gangrene may result. Doctors unfamiliar with this type of injury should reference a knowledgeable medical source.
• To avoid the risk of injury, do not use bare hands to check for hydraulic leaks.
• Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment
like Power-PRO. Install and place Power-PRO into service according to the EMC information in this manual.
Portable and mobile RF communications equipment can affect the function of Power-PRO.
• The use of accessories, transducers and cables other than those specified, with the exception of transducers
and cables sold by Stryker as replacement parts for internal components may result in increased emissions or
decreased immunity of the Power-PRO cot.
• The Power-LOAD system and the Power-PRO cot should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, observe the Power-PRO cot to verify normal operation in the configuration
in which it will be used.
• Power-PRO operates at the following frequencies: 70 - 125 kHz for inductive charging and 13.56 MHz±7 kHz,
Amplitude Modulated (OOK), ERP: -79.57 dBm. The Power-PRO cot may be interfered with by other equipment,
even if that other equipment complies with CISPR emission requirements.
• To minimize the potential of a cot tip occurring, the stacked (6516-127-000) and side-by-side (6516-128-000)
configurations should not be exposed to inclines greater than five degrees when in positions other than the low
height (transport) position.
• To minimize the potential of a cot tip occurring, the side-by-side (6516-128-000) configuration should not be exposed
to inclines greater than ten degrees when in the low height (transport) position.
Return To Table of Contents
1-14
Summary of Safety Precautions
Incubator adaptor (6516-129-000) is designed to secure only Drager
®
(or Air-Shields
6516-109-005 REV B
®
incubators to the 6510
®
Series) incubators.
www.stryker.com