HEADSECTION REPLACEMENT

Tools Required:
• 7/16" Combination Wrench
• 3/16" Hex Wrench
Procedure:
1.
Raise the cot to the full upright position.
2.
Using a 7/16" combination wrench and a 3/16" hex
wrench, remove the two screws (A) that secure the
cap bearings to the base litter interface bracket
(one on each side) (Figure 47).
3.
Squeeze the head release handles and slowly
remove the head section assembly.
4.
Reverse steps to reinstall.
5.
Verify proper operation of the unit before returning
it to service.