Checking The Cot Battery Power Level - Stryker Power-PRO IT Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CHECKING THE COT BATTERY POWER LEVEL

English
To check the battery power level, press the retract (–)
button (A) as shown in Figure 18 on the cot control
switch to activate the cot battery LED indicator (B) as
shown in "Figure 19: Foot End Control Enclosure" on
page
1-29.
The cot battery LED indicator is located at the
Power-PRO foot end control enclosure (shown as a
battery symbol).
• The LED is solid green when the battery is fully
charged or has adequately charged battery power.
• The LED flashes amber when the battery needs
to be recharged or replaced.
• The LED is solid amber to indicate a battery error.
See
the
SMRT™
Maintenance Manual for additional SMRT™ Pak and
SMRT™ Charger information.
Notes:
• Automatic charging will only occur with SMRT™ Pak batteries.
• Only use Stryker-approved batteries with Power-PRO.
• If applicable, Power-LOAD automatically charges the Power-PRO SMRT™ Pak battery when the cot is locked into
Power-LOAD in the transport position (no cable or connectors required). The cot battery LED indicator momentarily
flashes green to signify that it is charging.
WARNING
• To avoid risk of electric shock, never attempt to open the battery pack for any reason. If the battery pack case
is cracked or damaged, do not insert it into the charger. Return damaged battery packs to a service center for
recycling.
• Do not remove the battery when the cot is activated.
• Avoid direct contact with a wet battery or battery enclosure. Contact may cause injury to the patient or operator.
CAUTION
• Only use the battery and charger as specified in the SMRT™ Power System Operations/Maintenance Manual.
• The cot is not for use with an AC adapter.
• When charging a battery in an ambulance, locate the charger in an enclosed cabinet and out of patient reach
during transport.
• Ensure that the battery is fully charged prior to placing into service. An uncharged or depleted battery may cause
poor cot performance.
Return To Table of Contents
1-28
Power-PRO Cot User Controls
Power
System
Operations/
6516-109-005 REV B
A
Figure 18: Retract Button - Control Switch
www.stryker.com

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

6516

Table des Matières