POWER-PRO™ IT (CONTINUED)
English
The Model 6516 Power-PRO™ IT cot is intended for use in the electromagnetic environment specified below. The
customer or user of the Model 6516 Power-PRO™ IT cot should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
EN/IEC 61000-4-2
Electrostatic fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
EN/IEC 61000-4-8
Note: U
is the alternating current mains voltage prior to application of the test level.
T
Return To Table of Contents
1-94
EMC Information
Guidance and manufacturer's declaration - electromagnetic immunity
EN/IEC 60601
test level
+6 kV contact
+8 kV air
+2 kV for power
supply lines
+1 kV for input/
output lines
+8 kV differential
mode
+2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 seconds
3A/m
Compliance level
Electromagnetic environment - guidance
+6 kV contact
Floors should be wood, concrete or ceramic
+8 kV air
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Applies to:
• Cot
• SMRT™ Charger (6500-201-010)
+2 kV for power
Mains power quality should be that of a typical
supply lines
commercial or hospital environment.
Applies to:
+1 kV for input/
• SMRT™ Charger (6500-201-010)
output lines
+8 kV differential
Mains power quality should be that of a typical
mode
commercial or hospital environment.
Applies to:
+2 kV common
• SMRT™ Charger (6500-201-010)
mode
<5% U
Mains power quality should be that of a
T
(>95% dip in U
)
typical commercial or hospital environment.
T
for 0.5 cycle
If the user of the charger requires continued
40% U
operation during power main interruptions, it
T
(60% dip in U
)
is recommended that the device be powered
T
for 5 cycles
from an uninterrupted power supply or a
70% U
battery.
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 seconds
3A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
Applies to:
• Cot
• SMRT™ Charger (6500-201-010)
6516-109-005 REV B
www.stryker.com