Parameter
Pulse amplitude LV
Pulse width A
Pulse width RV
Pulse width LV
Sensitivity A
Sensitivity RV
Sensitivity LV
Refractory period A
Refractory period RV
Refractory period LV
Mode switching
Onset criterion
Resolution criterion
Intervention rate
Switches to
The basic rate with mode switching
Rate stabilization with
mode switching
PVARP
PVARP after PVC
Capture control A
Capture control RV
Capture control LV
Atrial overdrive pacing
Factory
Standard
Safe
SR DR HF
setting
program
3.0 V
3.0 V
—
0.4 ms
0.4 ms
—
0.4 ms
0.4 ms
1.0 ms
x
0.4 ms
0.4 ms
—
AUTO
AUTO
—
AUTO
AUTO
2.5 mV
x
AUTO
AUTO
—
AUTO
AUTO
—
250 ms
250 ms
300 ms
x
200 ms
200 ms
—
ON
ON
—
5-out-
5-out-of 8 —
of 8
5-out-
5-out-of 8 —
of 8
160 bpm 160 bpm
—
DDIR
DDIR
—
+10 bpm +10 bpm
—
OFF
OFF
—
AUTO
225 ms
—
(Start
250 ms)
400 ms
Automati-
—
cally set
ON
ON
OFF
x
ON
ON
OFF
ON
ON
OFF
OFF
OFF
—
Parameter
x
Vp suppression
x
x
IEGM recording (HAR)
x
x
IEGM recording (HVR)
x
Home Monitoring
x
x
Tolerances of Parameter Values
x
x
x
Parameter
x
x
Basic rate
x
x
x
Basic interval
x
x
Magnet rate (magnet interval)
x
x
Pulse amplitude
x
x
Pulse width
x
x
Sensitivity A
EN 45502-2-1 triangle pulse
x
x
x
x
Sensitivity RV/LV
EN 45502-2-1 triangle pulse
x
x
Refractory period
x
x
Maximum activity rate
Lead impedance
x
x
x
x
x
x
x
x
x
22
Factory
Standard
Safe
setting
program
OFF
OFF
—
ON
AT
OFF
ON
ON
OFF
OFF
OFF
OFF
Range of values
Tolerance
30 ... (5) ... 100 ... (10)
± 20 ms
... 200 bpm
1000 ms
± 20 ms
90 bpm (664 ms)
± 20 ms
0.2 ... 7.5 V
The greater value of ±50 mV
or +20/-25%
0.1 ... 1.5 ms
The greater value of ±20 µs
or ±10%
0.1 ... 0.2 mV
The greater value of ±0.1 mV
or ±20%
0.3 ... 7.5 mV
0.5 ... 7.5 mV
±20%
200 ... 500 ms
± 20 ms
80 ... 180 bpm
± 20 ms
100 ... 200 Ω
±50 Ω
201 ... 2500 Ω
±10%
SR DR HF
x
x
x
x
x
x
x
x
x
x