Tolerances Of Parameter Values - BIOTRONIK ProMRI Eluna 8 SR-T Manuel Technique

Stimulateur cardiaque, traitement de la bradyarythmie, traitement par resynchronisation cardiaque
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Parameter
Factory
settings
Refractory period A
AUTO
Refractory period RV
250 ms
Refractory period LV
200 ms
Mode switching
ON
Onset criterion
5-out-of 8
Resolution criterion
5-out-of 8
Intervention rate
160 bpm
Switches to
DDIR
The basic rate with 
+10 bpm
mode switching
Rate stabilization with 
OFF
mode switching
PVARP
AUTO (Start
250 ms)
PVARP after PVC
400 ms
Capture control A
ON
Capture control RV
ON
Capture control LV
ON
Atrial overdrive pacing
OFF
Vp suppression
OFF
IEGM recording (HAR, HVR) ON
Home Monitoring
OFF
Standard
Safe
SR DR HF
program
program
AUTO
x
250 ms
300 ms
x
x
200 ms
200 ms
ON
x
5-out-of 8
x
5-out-of 8
x
160 bpm
x
DDIR
x
+10 bpm
x
OFF
x
AUTO (Start
x
250 ms)
400 ms
x
ON
OFF
x
x
ON
OFF
x
ON
OFF
OFF
x
OFF
ON
OFF
x
x
OFF
OFF
x
x

Tolerances of Parameter Values

Parameter
x
Basic rate
x
x
Basic interval
x
Magnet rate (magnet interval)
x
Pulse amplitude
x
x
Pulse width
x
x
Sensitivity A
x
EN 45502-2-1 triangle pulse
RV/LV sensitivity
x
EN 45502-2-1 triangle pulse
Refractory period
x
Maximum activity rate
x
Lead impedance
x
x
x
x
x
x
22
Range of values
Tolerance
30 ... (5) ... 100 ... (10)
± 20 ms
... 200 bpm
1000 ms
± 20 ms
90 bpm (664 ms)
± 20 ms
0.2 ... 7.5 V
The greater value of ±50 mV
or +20/-25%
0.1 ... 0.4 ms
± 0.04 ms
0.5 ... 1.0 ms
± 0.10 ms
1.25 ... 1.5 ms
± 0.15 ms
0.1 ... 0.2 mV
±0,05 mV
0.3 ... 7.5 mV
±20%
0.5 ... 7.5 mV
±50%
200 ... 500 ms
+ 10/-30 ms
80 ... 180 bpm
± 20 ms
100 ... 200 Ω
±50 Ω
201 ... 2500 Ω
±25%