Refractory periods
Parameter
Range of values
RV refractory period
200 ... (25) ... 500 ms
Atrial refractory
AUTO
period
Atrial refractory
300 ... (25) ... 775 ms
period in the modes
AAI(R); AAT(R); DDT
LV refractory period
200 ms
AUTO PVARP
OFF; ON
PVARP
175 ... (25) ... 600 ms
PVARP after PVC
PVARP + 150 ms
(max: 600 ms)
Blanking periods
Parameter
Range of values
Far-field protection
100 ... (10) ... 220 ms
after Vs
Far-field protection
100 ... (10) ... 220 ms
after Vp
Ventricular blanking
30 ... (5) ... 70 ms
period after Ap
PMT protection
Parameter
Range of values
PMT protection
OFF; ON
VA criterion
250 ... (25) ... 500 ms
Standard
SR DR HF
250 ms
x
x
x
AUTO
x
x
350 ms
x
x
200 ms
x
ON
x
x
225 ms
x
x
Automatically set
x
x
Standard
SR DR HF
100 ms
x
x
150 ms
x
x
30 ms
x
x
Standard
SR DR HF
ON
x
x
350 ms
x
x

Pacing and Sensing

Pulse amplitude and pulse width
Parameter
Range of values
Pulse amplitude
0.2 ... (0.2) ... 6.0 ... (0.5)
A/RV/LV
... 7.5 V
Pulse width A/RV/LV
0.1 ...(0.1) ... 0.5 ... (0.25)
... 1.5 ms
Sensitivity
Parameter
Range of values
Sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
Sensitivity A
OFF; AUTO; 0.1 ... (0.1) ... 1.5
... (0.5) ... 7.5 mV
RV sensitivity
AUTO; 0.5 ... (0.5) ... 7.5 mV
LV sensitivity
OFF; AUTO; 0.5 ... (0.5) ...
7.5 mV
Atrial capture control
Parameter
Range of values
Atrial capture
ATM (monitoring only); ON;
control
OFF
Minimum amplitude
0.5 ... (0.1) ... 4.8 V
Threshold test start
2.4 ... (0.6) ... 4.8 V
Safety margin
0.5 ... (0.1) ... 1.2 V
Search type
Interval; time of day
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
Time of day
00:00 ... (00:10) ...
23:50 hh:mm
18
Standard
SR DR HF
3.0 V
x
x
x
0.4 ms
x
x
x
Standard
SR DR HF
AUTO
x
AUTO
x
x
AUTO
x
x
x
AUTO
x
Standard
SR DR HF
ON
x
x
1.0 V
x
x
3.0 V
x
x
1.0 V
x
x
Time of day
x
x
24 h
x
x
00:30 hh:mm
x
x