AV hystereses
Parameter
Range of values
AV hysteresis mode
OFF; Positive; Negative
HF when setting RV: IRSplus
Positive modes:
70; 110; 150; 200 ms
AV hysteresis
Negative modes:
10 ... (10) ... 150 ms
AV hysteresis
AV repetetive /
OFF; ON
scan cyles
Ventricular pacing
Parameter
Range of values
Ventricular pacing
BiV, RV; LV
Triggering
OFF; RVs; RVs + PVC
LV T-wave protection
ON; OFF
Maximum trigger rate AUTO; 90 ... (10) ... 160 bpm
Initially paced
RV; LV
chamber
VV delay after Vp
0 ... (5) ... 80 ... (10) ... 100 ms 0 ms
VV delay after Vs
0 ms
Upper rate
Parameter
Range of values
Upper rate
90 ... (10) ... 200 bpm
SR: in VVT mode
Wenckebach response/
Automatically set
2:1 rate
Atrial upper rate
OFF; 175; 200; 240 bpm
en • English
Mode switching
Standard
SR DR HF
Parameter
OFF
x
x
Mode switching
Intervention rate
70 ms
x
x
Switch to mode
CLS modes:
110 ms
50 ms
x
x
Ventricular pacing
ON
x
x
Onset criterion
Resolution criterion
Change of the basic rate
with mode switching
Standard
SR DR HF
Rate stabilization with
BiV
x
mode switching
RVs
x
2:1 lock-in protection
ON
x
AUTO
x
LV
x
Ventricular pacing suppression
Parameters valid for devices in DDD-ADI or DDDR-ADIR modes:
x
Parameter
0 ms
x
Vp suppression
Pacing suppression after
consecutive Vs
Standard
SR DR HF
Pacing support after x cycles
130 bpm
x
x
x
x
x
240 bpm
x
x
17
Range of values
Standard
OFF; ON
ON
100 ... (10) ... 250 bpm
160 bpm
DDI; DDI(R) when
DDI(R)
permanent DDD(R)
VDI; VDI(R) when
permanent VDD(R)
RV; BiV
BiV
3 ... (1) ... 8 (out of 8)
5
OFF; +5 ... (5) ... +30 bpm +10 bpm
OFF; ON
OFF
OFF; ON
ON
When setting RV: OFF;
ON
ON
Range of values
Standard
OFF; ON
OFF
1 ... (1) ... 8
6
1 ... (1) ... 4 (out of 8) 3