Ventricular capture control
Parameter
Range of values
Capture control RV
ATM (monitoring only); ON;
OFF
Capture control LV
Minimum
0.7 V
amplitude RV
Minimum
amplitude LV
Threshold test start
2.4 ... (0.6) ... 4.8 V
RV safety margin
0.3 ... (0.1) ... 1.2 V
LV safety margin
1.0; 1.2 V
Search type
Interval; time of day
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
Time of day
00:00 ... (00:10)
... 23:50 hh:mm
Atrial overdrive pacing
Parameter
Range of values
Atrial overdrive pacing OFF; ON
With ON: maximum over-
pacing rate 120 bpm, mean
rate increase approximately
8 bpm, rate decrease after
20 cycles
en • English
Lead configuration
Standard
SR DR HF
Parameter
ON
x
x
x
Sensing polarity A
x
Sensing polarity RV
0.7 V
x
x
x
Sensing polarity LV
Pacing polarity A
x
Pacing polarity RV
Pacing polarity LV
3.0 V
x
x
x
0.5 V
x
x
1.0 V
x
Time of day
x
x
x
24 h
x
x
x
00:30 hh:mm x
x
x
Standard
SR DR HF
OFF
x
x
19
Range of values
Standard
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Unipolar; bipolar
Unipolar
Device type HF:
LV1 tip –>
LV1 tip -> LV2 ring
housing
LV1 tip -> RV ring
LV2 ring -> LV1 tip
LV2 ring -> RV ring
LV1 tip -> housing
LV2 ring -> housing
Device type HF QP
LV1 tip –>
LV1 tip -> LV2 ring
LV2 ring
LV1 tip -> LV4 ring
LV1 tip -> RV ring
LV1 tip -> housing
LV2 ring -> LV1 tip
LV2 ring -> LV4 ring
LV2 ring -> RV ring
LV2 ring -> housing
LV3 ring -> LV2 ring
LV3 ring -> LV4 ring
LV3 ring -> RV ring
LV4 ring -> LV2 ring
LV4 ring -> RV ring
SR DR HF
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x