• Telemetry information for the USA:
Telemetry data for the USA: This transmitter is authorized by rule under the
Medical Device Radiocommunication Service (in part 95 of the FCC Rules) and must
not cause harmful interference to stations operating in the 400.150-406.000 MHz
band in the Meteorological Aids (i.e., transmitters and receivers used to communi-
cate weather data), the Meteorological Satellite, or the Earth Exploration Satellite
Services and must accept interference that may be caused by such stations,
including interference that may cause undesired operation. This transmitter shall
be used only in accordance with the FCC Rules governing the Medical Device Radio-
communication Service. Analog and digital voice communications are prohibited.
Although this transmitter has been approved by the Federal Communications
Commission, there is no guarantee that it will not receive interference or that any
particular transmission from this transmitter will be free from interference.
This device will be registered with Federal Communications Commission under the
following number:
FCC ID: QRIPNP
Pulse form
The pacing pulse has the following form:
Resistance to interference
All variants of BIOTRONIK devices comply with the requirements of EN 45502-2-1:
2003, § 27.5.1 at the highest sensitivity.
The pulse amplitude reaches its maximum value
at the beginning of the pulse (Ua). With
increasing pacing duration (tb), the pulse
amplitude is reduced dependent on the pacing
impedance.
Battery Data
Battery characteristics
The following data is provided by the manufacturers:
Wilson
Manufacturer
GREATBATCH, INC.
Clarence, NY 14031
Battery type
GB 3193
System
QMR
Li-CFX/SVO
Device type
SR; DR
Battery voltage at BOS
3.3 V
Open-circuit voltage
3.3 V
Nominal capacity
1010 mAh
Usable capacity until EOS
971 mAh
Remaining capacity at ERI
39 mAh
Shortening of the service time after long storage period
In case of implantation after an average storage period – about 1 year before the end of
the use by date – the average service time decreases by about 1%.
Devices should be implanted within 19 months between the manufacturing date and the
use by date (indicated on the package).
Power consumption
• BOS, inhibited: SR(-T), DR(-T) 6 µA; HF-T (QP) 7 µA
• BOS, 100% pacing: SR(-T) 8 µA; DR(-T) 11 µA; HF-T (QP) 14 µA
Calculation of service times
Mean service times pre-estimated from the following and other data:
• Storage for 6 months
• Technical data of the battery manufacturer
• Basic rate of 60 bpm in AAIR/VVIR modes (single-chamber devices) or DDDR modes
(dual-chamber and triple-chamber devices)
• Home Monitoring configuration: OFF
• No wandless telemetry
• Configuration of different pulse amplitudes and lead impedances
24
LITRONIK GmbH
01796 Pirna
Germany
LiS 2650MK
®
LiMn0
2
SR; DR
3.1 V
3.1 V
950 mAh
880 mAh
70 mAh
LiS 3150MK
LiMn0
2
HF; HF QP
3.1 V
3.1 V
1200 mAh
1066 mAh
134 mAh