IEGM recordings
Parameter
Range of values
Number of recordings
6 series: 12
(each max. 10 s)
8 series: 20
High atrial rate (HAR) OFF; AT; mode switching
High ventricular rate
OFF; ON
(HVR)
8 series:
OFF; ON
Patient triggering
(triggered by patient)
Pre-trigger recording 0; 25; 50; 75; 100%
IEGM signal
Filtered; Unfiltered
Rates for statistics
Parameter
Range of values
HAR limit
100 ... (10) ... 250 bpm
HVR limit
150 ... (5) ... 200 bpm
HVR counter
4; 8; 12; 16 events
Start resting period
00:00 ... (1:00) ... 23:00 hh:mm 2:00 hh:mm
Duration of resting
0.5 ... (0.5) ... 12 h
period
Lead check
OFF; ON

Rate Adaptation

CLS modes: closed loop stimulation
Standard
SR DR HF
Parameters valid for 8 series devices:
x
x
x
Parameter
Maximum CLS rate
AT
x
x
x
CLS response
ON
x
x
x
CLS resting rate
OFF
x
x
x
control
Vp required
75%
x
x
x
Filtered
x
x
x
R modes: Accelerometer
Parameters valid for devices with R modes:
Standard
SR DR HF
Parameter
200 bpm
x
x
Sensor gain
180 bpm
x
x
x
8 events
x
x
x
Max. activity rate
x
x
x
Sensor threshold
4 h
x
x
x
Rate fading
ON
x
x
x
Rate increase
Rate decrease
20
Range of values
Standard
80 ... (10) ... 160 bpm
120 bpm
Very low; Low; Medium; High;
Medium
Very high
OFF; +10 ... (10) ... +50 bpm
+20 bpm
Yes; No
No
When BiV is
set: Yes
Range of values
Standard
AUTO; Very low; Low;
AUTO
Medium; High; Very high
80 ... (10) ... 180 bpm
120 bpm
Very low; Low; Medium; High;
Medium
Very high
OFF; ON
OFF
1; 2; 4; 8 bpm/cycle
2 bpm/cycle
0.1; 0.2; 0.5; 1.0 bpm/cycle
0.5 bpm/
cycle
SR DR HF
x
x
x
x
x
x
x
x
x
x
x
x
SR DR HF
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x