2. Start the program.
3. Remove the instruments from the disinfection unit after the program
has fi nished.
4. Inspect and pack the instruments immediately after removal (after
additional post-drying in a clean place).
Do not blow the air from the distal end to the proximal end!
3. Visual inspection and function testing
DE
Check the devices..
ß Inspecting devices, page 17
EN
4. Packaging
Prior to sterilization, the disassembled instruments must be packaged in sui-
FR
table disposable sterilization packaging (single or double pouch) and/or a
suitable sterilization container:
ES
IT
PL
22
Step 3:
Because fl uid penetrates the gas feed line of the connec-
tion cable as well as the probe during cleaning, blow out
and dry the relevant adapters using fi ltered compressed
air at a pressure of less than 3 bar after cleaning. To do
this, place the compressed air attachment loosely on the
Luer lock connector of the connection cable/probe.
in accordance with DIN EN (ANSI AAMI) ISO 11607/
DIN EN 868-2...10 (previously DIN EN 868/ANSI AAMI
ISO 11607)
Suitable for steam sterilization (temperature resistance
up to 137 °C, suffi cient steam permeability)
serviced regularly (sterilization container)
Sterilization in the shipping packaging (e.g. silicone protective
cap, etc.) is prohibited!