Operating Notes
Operating Notes
General Remarks
Caution!
The connections of Reocor D and the temporary
pacing leads must be secured and checked regularly.
After the device is switched on, Reocor D executes a
Self-test
self-test for a few seconds. This includes:
• Check of the program code and the microproces-
sor
• Memory test
• Function test of the LEDs and the acoustic signals
• Test of the pacing and sensing capability
• Test of the efficacy of high rate protection
When the self-test finds a defect, all LEDs flash con-
tinuously and acoustic warning signals sound. In this
case, the pacemaker must be turned off and sent to
BIOTRONIK.
If the self-test did not find any errors, Reocor D begins
delivering pacing pulses in accordance with the
programmed parameters. The negative electrode
(cathode) should therefore only be connected when
it has been ensured that the pacing mode, pacing
rate, pulse amplitude and sensitivity have been pro-
grammed correctly.
Setting the rotary switch for the operating mode to
OFF prevents pacing pulses from being delivered to
the patient immediately after connecting the leads.