Checking the
Test level
resistance to
according to
interference
IEC 60601
Conducted HF
10 V
interferences
10 kHz to 80 MHz
according to
outside of the ISM
IEC 61000-4-6
bands
10 V
10 kHz to 80 MHz
inside of the ISM
bands
Radiated HF
10 V/m
interference
800 MHz to 2.5 GHz
according to
IEC 61000-4-3
Conformity According to IEC 60601-1-2
Manufacturer guidelines and declaration –
resistance to electromagnetic interference for
all external pacemaker models 
(IEC 60601-1-2: Table 3)
The device is intended for use in an electromagnetic
environment as described below. The user of the
device should make sure that it is used in such an
environment.
Compliance
level
10 V
rms
a
10 V
rms
a)
10 V/m
Guidelines for the
electromagnetic environment
Portable and mobile radio
devices are not used closer to
any part of the device, includ-
ing cables, than the recom-
mended safe distance.
Recommended safe distance:
rms
d = 0.35
rms
d = 1.2
d = 1.2
for 80 MHz to 800 MHz
d = 2.3
for 800 MHz to 2.5 GHz
P
P
P
P