Operating Notes
Connection
WARNING!
Danger to patient by damaged cables.
Damaged cables are limited in functionality and pose
a danger to patients. Do not use damaged cables.
WARNING!
Danger from loss of function.
Damp cables have limited functionality and pose a
danger to patients. Do not use damp cables.
WARNING!
Danger from electrical currents.
Unused cable contacts can conduct electrical currents
to patients. Adhere unused cable contacts close to the
patient.
Caution!
Allergic reactions and inflammations.
Prevent the cable from coming into contact with the
patient's wounds or skin.
Note:
Ensure correct fitting of the insulators prior to using
the cables.
Note:
Do not connect the patient cable to the temporary
pacing lead of the patient before the connection has
been established to the Reocor D.
Direct connection
If Reocor D is used without the Redel adapter, tempo-
rary catheters and heart wires can be connected to the
patient cables PK-82 and PK-83 directly at the con-
nector ports A+, A- and V+, V-.
Patient cable
The patient cable is connected via a Redel adapter to
the Reocor D.
Note:
No dual-chamber pacing is possible with the patient
cable PK-83-B. Leads that are connected with the
PK-83-B are connected to the ventricular channel of
Reocor D.
Fit the Redel adapter to Reocor D.
Screw the adapter in tight.
Insert the Redel plug of the patient cable into the
Redel port of the adapter.