Connection:
BIOTRONIK cable
patient-side
IS-1 connector
PK-67-S/L with PA-2
port
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155 (2x) Redel plug
Alligator clips
PK-82 (2x)
Operating Notes
Implanted lead with IS-1 connector
Recommended connections
Possible connections
Polarity
Reocor D principally paces in bipolar mode, but it can
be used with bipolar or unipolar temporary pacing
leads.
If unipolar leads are used, two leads must be con-
nected for each chamber.
Separating connections
Disconnect patient cables from the temporary pacing
leads of the patient or disengage the direct connec-
tion.
Separating Redel plug
• Retract the retaining ring at the Redel plug and
pull the Redel plug off the Redel port.

Start Up

The operation of Reocor D is identical for all operating
modes. The operating steps should be carried out in
the following order (the numbers in parentheses refer
to Figure 1 on page 64).
• Insert battery.
• Push protective cover down.
• Prepare patient: Place the leads but do not
connect them to the pacemaker yet.
Connection
device-side
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel adapter
2-mm plug
2-mm connector port
Reocor D
connection