General Safety Instructions; Operating Conditions - BIOTRONIK ProMRI Eluna 8 SR-T Manuel Technique

Stimulateur cardiaque, traitement de la bradyarythmie, traitement par resynchronisation cardiaque
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Resynchronisation therapy
Triple chamber devices have functions to configure different VV delays in order to
resynchronize the ventricles.
• Capture Control is available for the left ventricle with automated tracking of the
pacing threshold or automatic threshold monitoring (ATM) for trend analysis.
• To ensure that no additional surgery is necessary in case of a left-sided increase of
pacing threshold or undesired phrenic nerve stimulation, different pacing polarities
can be set for the left ventricular lead with a triple-chamber device.
• An additional diagnostic function with biventricular pacing: variability of the heart
rate, the patient activity and the thoracic impedance are monitored on a continual
basis.
Programs
There are two types of therapy programs:
• Default parameters are offered for the most common indications (Program Consult
function).
• Individual settings can be saved in 3 individual therapy programs.
Home Monitoring functions
The device automatically sends information to the transmitter once a day. Additionally,
the test messages can be initiated using the programmer. Important medical informa-
tion includes, among others, the following:
• Ongoing atrial and ventricular arrhythmia
• Parameters relevant to leads in the atrium and ventricle: thresholds, sensing
amplitudes, impedances
• Current statistics on bradycardia therapy
• Individually adjustable timing interval for device messages which provide additional
information pertaining to the device messages
®
• IEGM online HD
with up to 3 high definition channels
• Transmission of these IEGM recordings along with device messages
en • English

2 General Safety Instructions

Operating Conditions

Technical manuals
The following technical manuals provide information about usage of the device
systems:
— Technical manual for the device
— Technical manual for the HMSC
— Technical manuals for the programmer and its accessories
— Technical manuals for the user interface
— Technical manuals for cables, adapters and accessories
• Technical manuals are either included in hard copy form in the storage package or
in digital form on the internet:
manuals.biotronik.com
• Follow all relevant technical manuals.
• Preserve technical manuals for later use.
Care during shipping and storage
• Devices are not to be stored close to magnets or sources of electromagnetic inter-
ference.
• Note the effects of the storage duration; see Battery Data.
Temperature
Extremely low and high temperatures affect the service time of the battery in the
device.
• Permitted for shipping and storage:
–10°C to 45°C
Sterile delivery
The device and the screwdriver have been gas-sterilized. Sterility is guaranteed only if
the blister and quality control seal have not been damaged.
Sterile packaging
The device and screwdriver are each packaged in 2 separately sealed blisters. The
inner blister is also sterile on the outside so that it can be transferred in a sterile state
during implantation.