65
Designation
1a Atrial channel connection For cables and leads with 2-mm plugs
1b Ventricular channel
connection
2 Battery compartment
3 Low battery LED
4 Ventricle Pace LED
5 Ventricle Sense LED
6 Ventricle Ampl. control
dial
7 Ventricle Sens. control
dial
8 Atrium Pace LED
9 Atrium Sense LED
10 Atrium Ampl. control dial Setting the atrial pulse amplitude
11 Atrium Sens. control dial Setting the atrial sensitivity
12 Start burst
13 Velcro harness and
hanger (on back)
14 Burst rate atrium control
dial
15 AV delay control dial
16 Select burst
17 Mode dial
18 Rate control dial
Table 1: Description of elements in Figure 1
Operating Notes
Function
or for Redel adapters
(red = plus; blue = minus)
For 9-V block battery
Provides a warning when the battery voltage is too low
Yellow display for ventricular stimulated event
Green display for sensed R wave
Setting the ventricular pulse amplitude
Setting the ventricular sensitivity
(cannot be used in D00 and V00 operating modes)
Yellow display for atrial stimulated event
Green display for sensed P wave
(cannot be used in VDD operating mode)
(cannot be used in D00 operating mode)
Starts the atrial burst function
Securing Reocor D to patient, bed or infusion stand
Setting the atrial burst rate
Setting the AV delay
Selection of the atrial burst function
Selection of the pacing operating mode and off switch
Setting the pacing rate
Bold labels of the operating devices indicate safe
values for the intended use of the device.