Maintenance, Service, Inspections - BIOTRONIK CRM Reocor D Manuel Technique

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 148
84
Handling, Care and Maintenance

Maintenance, Service, Inspections

The only required maintenance action is the replace-
ment of the battery (see page 75).
No other maintenance work is required.
A short test should be performed prior to each use 
Test before use
of the device. It consists of a visual inspection and a
simple function test.
Visual inspection:
• Inspect the housing for mechanical damage,
• Inspect the cable connection area for mechanical
• Inspect the labeling for legibility.
Function test:
The result of the self-test that runs automatically
after activation must be heeded.
Inspections should be performed:
Inspection
• after an application together with high-frequency
• when malfunctions are suspected,
• once a year.
The inspection should follow the manufacturer speci-
fications. These are made available upon request.
The specification lists all required test steps and the
necessary equipment.
deformation, loose parts, cracks, etc.
damage.
surgical instruments or defibrillators,