Av Delay; Pulse Amplitude - Atrium/Ventricle; Sensitivity - Atrium; Sensitivity - Ventricle - BIOTRONIK CRM Reocor D Manuel Technique

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 148
79
Pacing Modes and Parameters

AV Delay

The AV delay can be continuously adjusted from 15 ms
to 400 ms with the AV delay control dial (15). Short AV
delays may be selected for special indications, e.g., in
case of recurring tachycardia.
Reocor D limits the AV delay upwards to half the basic
rate interval.
Pulse Amplitude – Atrium/Ventricle
The pulse amplitudes for the atrium and ventricle can
be adjusted with the Ampl. control dials (10), (6) within
a range from 0.1 V to 17 V. If a value less than 1 V is
set, the device will sound a warning signal for
two seconds.
The pulse width is 1 ms.
Pacing should be checked regularly to ensure that
pacing takes place and that a sufficient safety margin
has been set.
Sensitivity – Atrium
The sensitivity can be adjusted with the Atrium Sens.
control dial (11) between 0.2 mV and 10 mV. It should
be checked regularly to ensure that correct sensing
takes place and that a sufficient safety margin has
been set.
Sensitivity – Ventricle
The sensitivity can be adjusted with the Ventricle
Sens. control dial (7) between 1 mV and 20 mV. The
sensitivity should be checked regularly to ensure that
correct sensing takes place and that a sufficient safety
margin has been set.

Publicité

Table des Matières
loading

Table des Matières