72
Connection:
BIOTRONIK cable
patient-side
Direct connection (without BIOTRONIK cables)
2 mm
PK-67-S/L with PA-1-B
Screw terminals PK-175 with TC Adapt
Screw terminals PK-83 with TC Adapt (2x)
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155 (2x) Redel plug
Alligator clips
PK-82 (2x)
Connection:
BIOTRONIK cable
patient-side
Screw terminals PK-175
Screw terminals PK-83 (2x)
Alligator clips
PK-141
Alligator clips
PK-67-S/L with PA-4
Alligator clips
PK-67-S/L with PK-155 (2x) Redel plug
Alligator clips
PK-82 (2x)
Operating Notes
Connection Variants
Temporary catheters with 2-mm plugs or
heart wire with 2-mm adapter
You have the option to connect Reocor D directly to a
temporary catheter with touch-proof 2-mm plug or to
a heart wire with 2-mm adapter, without any other
cables or adapters. All other options are listed in the
following table.
Recommended connections
Possible connections
Heart wire with break-off needle or with
flexible end (max. 2.3 mm in diameter)
Recommended connections
Possible connections
Connection
device-side
2-mm connector port
Redel plug
Redel adapter
Redel plug
Redel adapter
2-mm plug
2-mm connector port
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel adapter
2-mm plug
2-mm connector port
Connection
device-side
Redel plug
Redel adapter
2-mm plug
2-mm connector port
Redel plug
Redel adapter
Redel plug
Redel adapter
Redel adapter
2-mm plug
2-mm connector port
Reocor D
connection
Reocor D
connection