Visual And Acoustic Signals - BIOTRONIK CRM Reocor D Manuel Technique

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 148
61
Warning
Acoustic signal for 2 s A pulse amplitude of
Fast sequence of
sounds
Acoustic signal and
flashing of the Pace
and Sense LEDs
Low battery LED
flashes.
a)
When using the battery type MN 1604 Duracell
General Description

Visual and Acoustic Signals

• During the self-test after switching on Reocor D,
all LEDs light up and brief acoustic signals can 
be heard. The self-test is completed after a few
seconds.
• If the self-test does not find any errors, the LEDs
and warning signals turn off.
• When the self-test finds a defect, all LEDs flash
continuously and warning signals sound.
• A required battery replacement is indicated by the
flashing red Low battery LED.
• The Sense (green) LEDs signal sensing of a P wave
or R wave.
• The Pace (yellow) LEDs signal pulse delivery.
• The LEDs and acoustic signals also provide the
following warnings during operation:
Meaning
or a rate of
180 ppm is
>
programmed.
Impedance outside of the
permissible range
High rate protection has
been triggered;
self-test failed.
ERI has been reached.
Error correction
1 V
Check whether the set
<
values are suitable for
the patient.
Check whether all
connectors are
securely plugged in.
Check whether the
leads have the desired
position.
Turn the device off and
return it to BIOTRONIK.
Replace the battery;
about 36 hours
service time remain.
®
®
, Procell
a)
of