74
Caution!
1
Only for 2-chamber pacing
Operating Notes
• Prepare Reocor D:
Set the pacing rate with the Rate control dial (18).
Set the AV delay with the AV delay control dial (15)
Set the pacing amplitudes for atrium
cle with the Atrium Ampl. (10)
(6) control dials.
• Select the pacing mode with the dial Mode (17).
The device will be activated at the same time.
• After the internal self-test has been completed
successfully, all 5 LEDs on the operating panel will
simultaneously flash twice.
• If the Low battery LED (3) flashes, the battery
needs to be exchanged (for battery exchange, 
see page 75).
• Connect leads, the yellow Atrium Pace (8)
Ventricle Pace (4) LEDs will flash in synchrony with
the atrial and ventricular pacing pulses.
• Set the sensitivity for atrium and ventricle with the
Atrium Sens. (11)
dials (7) such that the green Atrium Sense (9)
Ventricle Sense (5) LEDs flash in synchrony with
each sensed atrial or ventricular event.
A sufficient safety margin should be considered to
ensure reliable sensing.
• Monitor the ECG of the patient and adjust ampli-
tude and/or sensitivity, if necessary.
During use of Reocor D, the heart rate of the patient
must be monitored with an ECG monitor with an alarm
function.
1
and Ventricle Ampl.
1
and Ventricle Sens. control
1
.
1
and ventri-
1
and
1
and