Warning messages
Operating Notes
The following warnings can appear during use:
• A required battery replacement is indicated by the
flashing Low battery LED.
• If the lead impedance is not within a permissible
range (e.g. due to a fractured lead or a loose 
contact), a rapid sequence of sounds can be heard
no earlier than 5 seconds after activation.
• If the pulse amplitude is set to values
rate to values
for about 2 seconds.
• If the rate is too high (see page 88 "High rate
protection") or if the self-test has not passed, 
an acoustic signal sounds and the Pace and Sense
LEDs flash.
180 ppm, an acoustic signal sounds