• The three nebulizer parts (excluding tubing) may be disinfected in the microwave in a steam
cleaning bag (i.e. Quick-Clean
use. Remove from bag and allow to cool.
• Dry according to instructions in Section 3.1.
E. ELECTRONIC STEAM STERILIZER
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.
• The three nebulizer parts (excluding tubing) may be disinfected using an electronic steam
sterilizer for baby bottles, following the manufacturer's instructions for use (e.g. AVENT
Electronic Sterilizer). The disinfection cycle should last approximately 15 minutes. Allow parts
to cool.
• Dry according to instructions in Section 3.1.
F. VINEGAR (Not recommended for high level disinfection)
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.
• The three nebulizer parts (excluding tubing) may be soaked in 1:3 vinegar to water solution
(approximately 1 cup vinegar to 3 cups water) for 1 hour. Rinse well with clean water.
• Dry according to instructions in Section 3.1.
3.1 DRYING
• To dry, shake the nebulizer parts to remove any excess water.
• We recommend you briefly reconnect the EZ Twist
and the compressed air supply to clear the air pathway.
• Allow parts to air dry thoroughly before reassembling.
• The nebulizer components may also be dried with a clean lint-free cloth prior to reassembly.
4. STORAGE
NEBULIZER
• Reattach the
NEBULIZER CUP
of the
MOUTHPIECE
• Reattach the
• Nebulizer components should be stored together (with tubing) in a dry, dust-free place avoiding
continuous exposure to direct sunlight.
5. HOSPITAL INSTRUCTIONS (For Multiple Patient Re-Use)
If the nebulizer will be used for multiple patients, it must be cleaned, disinfected and sterilized
between patients and before first use. The EZ Twist
replaced between patients.
The following are the cleaning and disinfectant procedures exclusively for qualified medical
specialists familiar with the hygiene requirements in hospitals and medical practices validated by
Trudell Medical International according to DIN EN 17664. If different sterilization procedures are
used, particular care should be taken to ensure they are suitable for cleaning or disinfecting medical
products and that they satisfy the requirements for material resistance.
†
†
MicroSteam
bag) following the manufacturer's instructions for
TOP, making sure to line up the
and gently hand tighten until a stop is detected.
to the device with the
TUBING
to the bottom of the nebulizer cup
WHITE CYLINDER
EXHALATION VALVE
TUBING
cannot be reprocessed and must be
†
STEM
with the
facing down.
7
EN