CAUTION - Ensure nebulizer components are allowed to air dry thoroughly between cleaning
steps. Condensation or residual wetness can present an increased risk through the growth of
bacteria.
5.1 CLEANING/DISINFECTION
CAUTION - Clean and disinfect immediately after use with a single patient. Clean, disinfect and
sterilize before every change in patient.
5.1.1 RECOMMENDED PROCEDURE: MECHANICAL CLEANING AND
DISINFECTION
• Prepare the nebulizer as described in Section 1 'Preparation'.
• Place the nebulizer components (excluding tubing) in the washer disinfector in a manner that
they can be cleaned most effectively.
• Select the Vario TD program.
• Following thermal disinfection, remove nebulizer components from washer disinfector and
shake to remove excess water.
• Allow to air dry thoroughly prior to continuing to sterilization.
• The effectiveness of this procedure has been proven using a Miele mechanical washer
disinfector with Neodisher
Weigert, Hamburg). The equipment used for thermal disinfection should conform to DIN EN ISO
15883.
5.1.2 RECOMMENDED PROCEDURE: MANUAL CLEANING AND DISINFECTION
CAUTION: The AEROECLIPSE* XL BAN (and its accessories) are not suitable for treatment
with disinfectants based on quaternary ammonium compounds. Quaternary ammonium compounds
tend to become concentrated in plastic materials, which can cause incompatibility reactions in
patients.
• Prepare the nebulizer as described in Section 1 'Preparation'.
• Place the nebulizer components (excluding tubing) in a solution of Korsolex
and approximately 50°C warm water and clean thoroughly with a brush. This process should
take approximately 5 minutes.
• Next, place the nebulizer components (excluding tubing) in a disinfectant solution of Korsolex
basic 4% for 15 minutes.
• Thoroughly rinse the nebulizer components under hygienically tested water, shake to remove
excess water and leave to air dry thoroughly on a clean, dry and absorbent surface.
• The effectiveness of this procedure has been proven using Korsolex
agent in combination with the Korsolex
5.2 STERILIZATION
Your nebulizer (excluding tubing) may be sterilized using the following validated procedures. If
another procedure is used, its effectiveness must be validated. Hospital hygiene standards should
be followed.
• A validated sterilization process can only be carried out on products once they have been
cleaned and disinfected following instructions in Section 5.1.
EN
8
Mediclean forte 0.5% and Neodisher
basic 4% disinfectant (Bode Chemie, Hamburg).
Z for neutralization (Dr.
EndoCleaner 0.5%
EndoCleaner 0.5% cleaning