EZ TWIST TUBING
Occasionally you may notice condensation on the inside of your tubing. This is normal. To
remove the condensation, simply connect one end of the EZ Twist
source, such as an Ombra* Compressor, and allow the air to run without the nebulizer attached.
This will force air through the tubing, drying the inside. To clean the exterior, wipe with a damp
cloth. Ensure the tubing is dry prior to storage.
3. DISINFECTION INSTRUCTIONS
The reusable AEROECLIPSE* XL BAN (excluding tubing) should be disinfected at least every other
treatment day using any of the methods listed below.
A. BLEACH
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.
• The three nebulizer parts (excluding tubing) may be soaked in 1:50 bleach to water solution
(approximately 1 tablespoon bleach in 3¼ cups water) for 3 minutes. Rinse well with clean water.
You may substitute a respiratory equipment disinfectant (such as Milton
for the bleach solution by following the manufacturer's instructions for use.
• Dry according to instructions in Section 3.1.
B. BOILING
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.
• The three nebulizer parts (excluding tubing) may be boiled in water for 15 minutes. Place the
individual nebulizer parts in the water once it has reached the boiling point. Remove from water
and allow to cool.
• Dry according to instructions in Section 3.1.
NOTE: For Cystic Fibrosis patients, guidelines recommend boiling for a full 10 minutes to
disinfect.
NOTE: Ensure there is enough water in the pan to avoid contact of the plastic parts with the
bottom of the hot pan.
C. ISOPROPYL ALCOHOL (70%)
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.
• Wipe the outside surfaces of the three nebulizer parts (excluding tubing) with a tissue dipped in
70% Isopropyl Alcohol then submerge the three parts in a soak of 70% Isopropyl Alcohol for 1
minute before shaking to remove any excess alcohol.
• Rinse well with clean water.
• Dry according to instructions in Section 3.1.
D. MICROWAVE STEAM BAG
• Prior to disinfection, follow Section 1 and 2 'Cleaning Your Nebulizer'. Do not dry or reassemble
the device.