Télécharger Imprimer la page

SICK LMS4000 Notice D'instruction page 78

Masquer les pouces Voir aussi pour LMS4000:

Publicité

12
ANNEXE
Table 29 : Telegram structure: sAN LMCstandby
Telegram structure: sAN LMCstandby
Telegram
Description
part
Command
Answer
type
Command Set device to standby
Status
Accepted when value is
code
0
Table 30 : Example: sAN LMCstandby
<STX>sAN{SPC}LMCstandby{SPC}0<ETX>
CoLa A
02
73 41 4E
20
4C 4D 43 73 74 61 6E 64 62 79
CoLa B 02 02 02 02 00 00 00 10
12.3.1.4.2.3
Start the laser and (unless in Standby mode) the motor of the device
Table 31 : Telegram structure: sMN LMCstartmeas
Telegram structure: sMN LMCstartmeas
(User level 'Authorized Client' required)
Telegram
Description
part
Command
Method
type
Command Start measurement
Table 32 : Example: sMN LMCstartmeas
<STX>sMN{SPC}LMCstartmeas<ETX>
<STX>sMN LMCstartmeas<ETX>
CoLa A
sMN LMCstartmeas
02 73 4D 4E 20 4C 4D 43 73 74 61 72 74 6D 65 61 73 03
02 02 02 02 00 00 00 10
CoLa B
73 4D 4E 20 4C 4D 43 73 74 61 72 74 6D 65 61 73
Table 33 : Telegram structure: sAN LMCstartmeas
Telegram structure: sAN LMCstartmeas
Telegram
Description
part
Command
Answer
type
Command Start measurement
78
N O T I C E D ' I N S T R U C T I O N | LMS4000
Variable
Length
String
3
String
10
Enum_8
1
73 41 4E
20
4C 4D 43 73 74 61 6E 64 62 79
Start measurement [sMN LMCstartmeas]
Variable
Length
String
3
String
12
73 4D 4E
20
4C 4D 43 73 74 61 72 74 6D 65 61 73
Variable
Length
String
3
String
12
Additional details
Success:
Not allowed:
20
30
03
20
00
Additional details
Additional details
Values CoLa A
Values CoLa B
(ASCII)
(Binary)
sAN
73 41 4E
LMCstandby
4C 4D 43 73
74 61 6E 64 62
79
0
00
1
01
49
Values CoLa A
Values CoLa B
(ASCII)
(Binary)
sMN
73 4D 4E
LMCstartmeas
4C 4D 43 73
74 61 72 74
6D 65 61 73
68
Values CoLa A
Values CoLa B
(ASCII)
(Binary)
sAN
73 41 4E
LMCstartmeas
4C 4D 43 73
74 61 72 74
6D 65 61 73
8027168/1MNR/2024-10-25 | SICK
Subject to change without notice

Publicité

loading