To flush the distal catheter:
• Occlude the catheter proximal to the reservoir with finger
pressure, then depress the reservoir.
• Failure of the reservoir to depress easily might indicate distal
catheter blockage.
Note: While proximal pressure is maintained, reservoir will not
refill with fluid.
Note: Flushing the distal catheter cannot be performed with
the right-angle valve configuration.
PRODUCT INFORMATION DISCLOSURE
INTEGRA LIFESCIENCES CORPORATION ("INTEGRA") HAS
EXERCISED REASONABLE CARE IN THE SELECTION OF
MATERIALS AND THE MANUFACTURE OF THESE PRODUCTS.
INTEGRA WARRANTS THAT THESE PRODUCTS SHALL
CONFORM TO THE PRODUCT LIMITED WARRANTY AS
PROVIDED IN THE PRODUCT LABELING OR APPLICABLE
PRODUCT CATALOG. THIS WARRANTY IS EXCLUSIVE, AND
INTEGRA DISCLAIMS ALL OTHER WARRANTIES, WHETHER
EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED
TO, ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. INTEGRA SHALL NOT
BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS,
DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING
FROM USE OF THESE PRODUCTS. INTEGRA NEITHER ASSUMES
NOR AUTHORIZES ANY PERSON TO ASSUME ANY OTHER OR
ADDITIONAL LIABILITY OR RESPONSIBILITY IN CONNECTION
WITH THESE PRODUCTS.
Appendix A
Optional Manometer Testing
Although Integra does not recommend functional testing, some
surgeons might choose to do so. Before testing, it is extremely
important that a CERTAS Programmable Valve with or without
SIPHONGUARD be flushed of all air bubbles. Air bubbles within
the valve or SIPHONGUARD will produce inaccurate manometer
test results. The presence of air bubbles can reduce the cross-
sectional area of the flow path, increase system resistance, and
impede the flow of fluid through the system during testing.
SIPHONGUARD Functional Testing
Equipment Required (use all sterile equipment, perform testing
under sterile conditions)
One sterile manometer, wide-bore (e.g. 3.5 mm), graduated in
mm (available in lengths from 38 to 60 cm)
One sterile 4-way stopcock
One sterile syringe, 10mL minimum is recommended
One sterile syringe filter, 5 µm
Sterile tubing adapters
Sterile silicone tubing
One sterile male luer connector with 1/16 in. (1.6 mm) barb
Sterile saline solution
Flushing Procedure
Note: At a rate of 0.5 mL/minute, unitized versions require
2–3 minutes to complete flushing. This is the time required for
fluid to fill the valve and exit the distal catheter. Allot additional
time to ensure the system is free of air bubbles.
1 . Fill syringe with sterile saline using the 5 µm syringe filter.
The syringe filter should not be reused in any subsequent
refilling of the syringe. Once the syringe has been filled,
remove the filter from the syringe.
2 . Assemble manometer, stopcock, syringe, and tubing
(Figure A-1).
3 . Adjust the valve setting to 1 while the valve remains in its
sterile package.