VIBRATION THERAPY
WARNING
• Always secure the support surface straps to the bed frame to prevent the support surface from sliding and to prevent
patient injury
• Risk of injury . Use of XPRT therapeutic support surface for stroke victims should be only under physician's order .
• To avoid possible injury, when using rotation or pulmonary therapy, take care not to extubate intubated patients .
• To ensure patient safety, always raise the bed siderails before beginning therapy .
• To avoid possible injury and to assure proper operation when using the XPRT™ therapeutic support surface on a
bed frame equipped with scales or bed exit follow manufacturer's instructions for use of these options .
• To avoid possible injury do not zero bed scales or weigh patient with Percussion, Vibration, Rotation or Turn
Assist active . Patient motion and position resulting from the support surface may adversely affect scale system
performance .
• To avoid possible injury do not initialize "Arm" bed exit with Percussion, Vibration, Rotation or Turn Assist active . The
patient motion and position resulting from the support surface may adversely affect bed exit system performance .

TO START VIBRATION THERAPY

1 .
Select the "Vibration" button .
2 .
Adjust the therapy parameters as needed, then
select the "Start" button .
a) Increase or decrease the duration of the
therapy .
b) Increase or decrease the intensity of the
vibration .
c) Select both lungs (bilateral) or either the left
of right lung to receive treatment .
www.stryker.com
Operation Guide