CPR ACTIVATION

WARNING
Deflate the XPRT™ therapeutic support surface or use
"Max Inflate" to inflate it completely before beginning
CPR . Refer to "To Start Max Inflate" on
1 .
To activate CPR, pull either the left, right or both
straps located at the head end of the support
surface . All running functions will stop and an alert
tone will sound .
2 .
If the left CPR strap is pulled, an alert tone will
sound to notify the caregiver and the display will
read "Left CPR Activated" until the plug is replaced
3 .
If the right CPR strap is pulled, an alert tone will
sound to notify the caregiver and the display will
read "Right CPR Activated" until the plug is replaced .
4 .
CPR will remain Activated until the plug is replaced .
5 .
If both CPR straps are pulled, an alert tone will sound
to notify the caregiver and the display will read "Left
CPR Activated" and "Right CPR activated" until the
plug is replaced .
6 .
To end CPR and resume therapy, replace the CPR
plugs until they snap back into their fully seated
position . When properly seated, the "CPR Activated"
on the Display will turn off .
www.stryker.com
Operation Guide
page 1-23
2950-109-005 REV B
Return To Table of Contents