Controller power supply circuitry is faulty, or the Controller received
a spike.
Press the START push-button twice to start the pump. Should the
message still be present, call the Agilent service.
S
Check connections and shortages between pump and controller, then
press the STOP RESET pushbutton twice to start the pump.
Orderable Parts
DESCRIPTION
Turbo-V Remote Panel
Remote Panel Connection Cable with valves remote
inputs and IP-54 protection grade
(Use with: TV2KG and TV3KT with on-board controller)
Remote Panel Connection Cable
(Use with: TV551, TV701 and TV1001 with navigator
controller)
EyeSys Mini-IMG Gauge cable (1 m)
EyeSys Mini-IMG Gauge cable (special)
EyeSys ConvecTor® Gauge cable (1 m)
Gauge extention cable (5 m)