17
Technical Information
Error Messages
The message will stay on until the temperature decreases below
threshold value. Press the STOP RESET pushbutton twice to start the
pump.
If the controller transformer temperature exceeds 90 °C, the pump is
shut off, and the display will be as shown in the following figure.
The message will stay on until the temperature decreases below
threshold value. Press the STOP RESET pushbutton twice to start the
pump.
Indicates a fault in the circuit of the controller used to power the
pump.
If the pump is stopped by an emergency stop signal provided via a
remote contact, the display will be as shown in the following figure.
Remove the controller power cable and check the emergency
condition.
Then reconnect the power cable and press the START pushbutton to
start the pump.
296/304