Turbo-V Remote Panel User Manual / 87-900-987-01
R1 (R2) TYPE: selects the type of reference measurement to
drive the relay. Possible types:
Frequency: driving frequency of the pump
Power: power supplied to the pump
Time: time from start
Normal: achievement of normal condition
Pressure: pressure read by the gauge, if present
R1 (R2) SET VALUE: Sets the reference value at which the relay
R1 (R2) is activated. This parameter is not available if SETPOINT
TYPE is set to NORMAL.
R1 (R2) HISTERESYS: Sets, in percentage, the hysteresis of the
relay. This parameter is not available if SETPOINT TYPE is set to
TIME or NORMAL.
R1 (R2) MASK TIME: sets the masked time (in hh:mm:ss),
measured from start, within which the SETPOINT R1 (R2) is not
driven even if the threshold value has been reached. This
parameter is not available if SETPOINT TYPE is set to TIME.
R1 (R2) OUTPUT LOGIC: Selects the control logic of the
SETPOINT output.
_|-- positive logic (the relay is energized when the measurement
exceeds the reference)
--|_ negative logic (the relay is energized when the measurement
is below the reference)
This parameter is not available if SETPOINT TYPE is set to TIME
or NORMAL.
Technical Information
(E)
17
Programming
289/304