MESSAGE
TOO HIGH L0AD
SHORT CIRCUIT
SYSTEM OVERRIDE
OVERVOLTAGE
POWER FAIL
Turbo-V Remote Panel User Manual / 87-900-987-01
DESCRIPTION
The current drawn by the pump is
higher than programmed.
After the starting phase the output
connection is shorted.
The pump is stopped by an emergency
stop signal provided via a remote
contact.
Controller power supply circuitry is
faulty, or the Controller received a
spike.
Failure in the controller's pump power
supply section.
Instructions for Use
Error Messages
REPAIR ACTION
Check that the pump rotor is free
to rotate.
Press the START push-button
twice to start the pump.
Check connections and shortages
between pump and controller.
Press the START push-button
twice to start the pump.
Check the emergency condition
which is causing the override
signal on the P1 connector.
Remove the override signal on P1
connector, then press the START
push-button twice to start the
pump.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent service.
Contact Agilent for Maintenance.