17
Technical Information
How to use by Front Panel
Start Stop and Reset Fail button
The front panel START/STOP function is available only if the controller is set to
NOTE
"FRONT mode". If it is set to "REMOTE mode", when the key is pressed an error
message is shown indicating "COMMAND LOCKED IN REMOTE", if it is set to
"SERIAL" mode, the "COMMAND LOCKED IN SERIAL MODE" error message is
displayed.
The "Reset Fail" function is always available regardless of controller mode.
Low Speed
The front panel LOW SPEED function is available only if the controller is set to
NOTE
"FRONT"; otherwise an error message is returned.
If it is set to "REMOTE" mode, the COMMAND LOCKED IN REMOTE error
message is displayed while if it is in "SERIAL" mode the "COMMAND LOCKED
IN SERIAL MODE" error message is returned.
268/304
Press once for the START command and again for the STOP
command. In the case of controller's failure, press once to reset
the fail condition.
Press once to activate the LOW SPEED command; press again to
de-activate.
Turbo-V Remote Panel User Manual / 87-900-987-01
(E)