All the logic output from the controller must be connected at P2
mating connector. The following table describes the signals available
on the connector.
Pins
Name
15-8
Fault
13-6
Starting / Running
12-5
Low Speed
11-4
R1
Turbo-V Remote Panel User Manual / 87-900-987-01
Description
0V = no fail
24V = fail
Configurable output.
Default setting (win.177 = 0):
24V
pump in RAMP state, 0V =
other state.
If win.177 = 1: 0V = pump in
stop or FAIL status, 24V =
other state
24V = low speed active
0V = low speed not active
Programmable Set Point Out
Can be related to:
Frequency, Power, Time
Status or Pressure.
See following figures for
more details. All the setting
can be done by: front panel
or serial line (win.101 to 105
and 162).
Technical Information
Controller Connection
Electrical
Optically isolated
Vmax (=open) = 24v
Vlow, max = 1,5v
Imax = 60mA
Pin-8 negative
Pin-15 positive
Optically isolated
Vmax (=open) = 24v
Vlow, max = 1,5v
Imax = 60mA
Pin-6 negative
Pin-13 positive
Optically isolated
Vmax (=open) = 24v
Vlow, max = 1,5v
Imax = 60mA
Pin-5 negative
Pin-12 positive
Optically isolated
Vmax (=open) = 24v
Vlow, max = 1,5v
Imax = 60mA
Pin-5 negative
Pin-12 positive
(E)
17
Note
Connect a 10kohm
resistor between
the 2 pins before
measuring voltage
A STOP command
(by front Panel or
serial or Start/stop
input) reset the
FAIL status.
Connect a 10kohm
resistor between
the 2 pins before
measuring voltage
Connect a 10kohm
resistor between
the 2 pins before
measuring voltage
Connect a 10kohm
resistor between
the 2 pins before
measuring voltage