17
Technical Information
Programming
282/304
CONTROLLER VENT: the valve is controlled by the controller
SERIAL COMMAND: the valve is controlled by the serial port.
VENT VALVE DELAY: Sets the delay, expressed in seconds,
between stopping of the pump and opening of the vent valve.
VENT VALVE OPEN TIME: sets the time, expressed in seconds,
of opening of the vent valve. Enter 0 to be keep the valve opened
indefinitely (until the next start).
LOAD GAS TYPE: selects the type of gas used in the process.
Therefore, the controller limits the power supplied to the pump
according to the setting of this parameter.
ACTIVE STOP: enables (YES) or disables (NO) active braking.
STOP SPEED READING: enables (YES) or disables (NO) the
pump rotational frequency reading function also during the
slowing phase that follows a stop command
Turbo-V Remote Panel User Manual / 87-900-987-01
(E)