17
Technical Information
Controller Connection
Pins
Name
14-7
R2
2-1
Programmable
Analog Output
9-1
Frequency
Analog Output
246/304
Description
Programmable Set Point Out
Can be related to:
Frequency, Power, Time
Status or Pressure.
See following figures for
more details. All the setting
can be done by: front panel
or serial line (win.171 to
176).
Programmable analog
output.
Can be related to :
frequency, power, pump
temperature (linear) or
pressure (logarithmic).
The setting can be done by:
front panel or serial line
(win.111).
0V = 0Hz, 0W, 0 °C
10V = 963Hz, 150W, 100 °C
See following chart for more
details
Analog output proportional
to frequency (linear).
0V = 0Hz
10V = 963Hz
Turbo-V Remote Panel User Manual / 87-900-987-01
Electrical
Optically isolated
Vmax (=open) = 24v
Vlow, max = 1,5v
Imax = 60mA
Pin-7 negative
Pin-14 positive
Optically isolated
Rload, min = 10kΩ
Ripple Vpp,max =
200mv
Precision = 5 %
Resolution = 40mv
Pin-1negative
Pin-2 positive
Optically isolated
Rload, min = 10kΩ
Ripple Vpp,max =
200mv
Precision = 1 %
Resolution = 40mv
Pin-1negative
Pin-9 positive
Note
Connect a 10kohm
resistor between
the 2 pins before
measuring voltage
(E)