Error Messages
Error Messages
For a certain type of failure, the Turbo-V Remote Panel will show a
list of error message described in the following table.
MESSAGE
COMMUNICATION ERROR
COMMUNICATION BROKEN
XXXXXXX MODEL
CHECK CONNECTION
TO PUMP
WAITING INTERLOCK
PUMP OVERTEMP.
CONTROLLER OVERTEMP.
234/304
DESCRIPTION
The Remote Panel is unable to start a
communication with the controller.
The Remote Panel is unable to
continue the communication with the
controller.
Controller not supported. The Remote
Panel works only with the following
pumps: TV551 with navigator
controller, TV701 with navigator
controller, TV1001 with navigator
controller, TV2k-G and TV3k-T.
Wrong connection between the pump
and the controller.
Or
The pump's temperature is below 0 °C.
The interlock signal of P1 connector is
activated by an interruption of the link
between pin 3 and 8 of J1 connector,
or because the external interlock signal
is open.
The upper bearing/pump temperature
exceeds 60 °C.
The controller environment
temperature exceeds 70 °C.
Or
The controller's radiator temperature is
above 60 °C.
Turbo-V Remote Panel User Manual / 87-900-987-01
REPAIR ACTION
Check serial connection between
the Remote Panel and the
controller.
Check serial connection between
the Remote Panel and the
controller.
Connect the Remote Panel to a
supported controller.
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
Reset the short circuit between
pin 3 and pin 8 of J1 connector,
or close the external interlock
signal.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice to start the pump.
Wait until the temperature
decrease below threshold value.
Press the START push-button
twice to start the pump.