17
Technical Information
Window Protocol Description
Window Meanings
N.
Read/ Write
000
R/W
001
R/W
107
R/W
101
R/W
103
R/W
104
R/W
105
R/W
108
R/W
262/304
Data
Description
Type
L
Start/Stop
(in remote/ Front mode the window is a read only)
L
Low Speed
Activation
N
Front
(default)/ Remote/ Serial configuration
N
R1 Set Point type
N
R1 Set Point value (expressed in Hz, W or s)
N
Set Point delay: time between the pump start and
the set point check (seconds)
N
Set point
hysteresis (in % of value)
N
Baud rate
Turbo-V Remote Panel User Manual / 87-900-987-01
Admitted
Values
Start = 1
Stop = 0
No = 0
Yes = 1
(default = 0)
Front = 0
Remote = 1
Serial = 2
(default = 0)
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 3)
0 to 99999
(default = 0)
0 to 100
(default = 2)
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
(E)