17
Technical Information
Window Protocol Description
N.
Read/ Write
176
R
200
R
203
R
205
R
206
R
206
R
207
R
208
R
221
R
224
R
225
R
408
R
264/304
Data
Description
Type
A
R2 Set Point Pressure Value
Valid in win 171 = 4
Format X.X E X.X
Where: X= 0 to 9
s = + or -
N
Pump
current in
mA dc
N
Driving frequency
in Hz
N
Pump status
N
Error code
N
Controller Heatsink Temperature ( °C)
L
R1 Set Point Status
L
R2 Set Point Status
L
R1 Set Point Status
A
Pressure reading
Format = X.X E XX
L
R2 Set Point Status
A
Firmware revision
Turbo-V Remote Panel User Manual / 87-900-987-01
Admitted
Values
Stop = 0
Waiting intlk = 1
Starting = 2
Normal = 3
High Load = 4
Fail = 5
Approaching low
speed = 6
Bit description:
see the following
figure
Off: 0
On: 1
Off: 0
On: 1
Off: 0
On: 1
Off: 0
On: 1
(E)

Publicité