Technical Information
Window Protocol Description
The controller can answers with the following response types:
Response Type
Response Length
Logic
1 byte
Numeric
6 bytes
Alphanumeric
10 bytes
ACK
1 byte
NACK
1 byte
Unknown Window
1 byte
Data Type Error
1 byte
Out of Range
1 byte
Win Disabled
1 byte
258/304
Response Value
-
-
-
(0x6)
(0x15)
(0x32)
(0x33)
(0x34)
(0x35)
Turbo-V Remote Panel User Manual / 87-900-987-01
Description
after a read instruction of a logic
window
after a read instruction of a numeric
window
after a read instruction of an
alphanumeric window
the command execution has been
successfully completed
the command execution has been
failed
the specified window in the command
is not a valid window
the data type specified in the command
(Logic, Numeric or Alphanumeric) is
not accorded with the specified
Window
the value expressed during a write
command is out of the range value of
the specified window
the specified window is Read Only or
temporarily disabled (for example you
can't write the Soft Start when the
Pump is running)
(E)