Télécharger Imprimer la page

Agilent Technologies Turbo-V Remote Panel Notice De Mode D'emploi page 243

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 42
Pins
Name
1-com
Start/Stop
2-com
Low Speed
3-com
Interlock
4-com
System Override
5-com
Soft Start
Turbo-V Remote Panel User Manual / 87-900-987-01
Description
Contact closed = Start
Contact open = Stop
Contact closed = Low Speed
Contact open = High Speed
Contact closed = Active
Contact open = Inactive
If active allows the pump running, if
inactive stops the pump.
Contact closed = Active
Contact open = Inactive
If inactive allows the pump running,
if active stops the pump and all the
others connected devices (Fan, Vent
Valve...).
Contact closed= Active
Contact open= Inactive
If active before the start command
activates the soft start procedure. To
be used only at first pump run-up.
Technical Information
Controller Connection
Electrical
Optically
isolated
Vlow, max =
10V
Imax = 5mA
Imin (on) = 2mA
Optically
isolated
Vlow, max =
10v
Imax = 5mA
Optically
isolated
Vlow, max =
10V
Imax = 5mA
Imin (on) = 2mA
Optically
isolated
Vlow, max =
10V
Imax = 5mA
Imin (on) = 2mA
Optically
isolated
Vlow, max =
10V
Imax = 5mA
Imin (on) = 2mA
(E)
17
Note
Active only in
Remote Mode
operation
Active only in
Remote Mode
operation
Active in all
modes
Active in all
modes
Active only in
Remote Mode
operation, and
in STOP status.
243/304

Publicité

loading

Ce manuel est également adapté pour:

969-8850