Technical Information
Programming
Gauge Menu
This menu contains the parameters relating to functioning of the
active gauge.
Figure 32
290/304
GAUGE
F. 1
Flow – 5/6
PRESSURE UNIT: selects the unit of measurement of the
pressure used to show this on the display, for display on the
serial line and for setting of the setpoint and autopressure
thresholds. Possible options:
mbar
Pascal
Torr
GAS CORRECTION: sets the conversion factor of the pressure
measurement in the case in which the process gas is not nitrogen.
The pressure indicated by the instrument is equal to:
P indicated = P measured x Gas correction / 10
Turbo-V Remote Panel User Manual / 87-900-987-01
PRESSURE UNIT
GAS CORRECTION
Flow 5/6
mbar
Pascal
Torr
01
02
99
(E)